Title |
A Simulation Study of Impedance Plethysmography for Diagnosing Deep Vein Thrombosis |
Authors |
이전(Lee, Jeon) ; 이경중(Lee, Kyoung-Joung) |
Keywords |
IPG ; VOP ; mathematical model ; hemodynamic change ; sensitivity coefficient ; electrode configuration |
Abstract |
In this study, the effects of vascular parameter changes and electrodes on VOP measurement based on IPG were simulated mathematically. For the evaluation of the effects of hemodynamic changes on VOP, a mathematical model, which consists of cardiovascular system model and venous occlusion model, was developed and the model solution representing the blood flow and pressure in measuring point was found by 2nd order Runge-Kutta method. And, with sensitivity coefficients obtained from finite element solution of electric field in measuring point, the effects of electrode system on measurement were evaluated. As increasing the resistance, the venous capacitance was not changed but the venous outflows were decreased and the decreased compliance reduced the venous capacitance. And, for several configurations of round electrodes and band electrodes, the sensitivity coefficients were computed using the electric field distribution along deep vein. In conclusion, the proposed mathematical cardiovascular model could be applied to the simulation study on the effects of hemodynamic parameters on DVT diagnosis with IPG. And, also the sensitivity coefficients could provide effective electrode configuration for exact measurement of VOP. |