Abstract |
This paper proposes a dry-type surface myoelectric sensor for the myoelectric hand prosthesis. The designed surface myoelectric sensor is composed of skin interface and processing circuits. The skin interface has one reference and two input electrodes, and the reference electrode is located in the center of two input electrodes. In this paper is proposed two types of sensors with the circle- and bar-shaped reference electrode, but all input electrodes are the bar-shaped. The metal material of the electrodes is the stainless steel (SUS440) that endures sweat and wet conditions. Considering the conduction velocity and the median frequency of the myoelectric signal, the inter-electrode distance (IED) between two input electrodes as 18mm, 20mm, and 22mm is selected. The signal processing circuit consists of a differential amplifier with a band pass filter, a band rejection filter for rejecting 60Hz power-line noise, amplifiers, and a mean absolute value(MAV) circuit. Using SUS440, six prototype skin interface with different reference electrode shape and IED is fabricated, and their output characteristics are evaluated by output signal obtained from the forearm of a healthy subject. The experimental results show that the skin interface with parallel bar shape and the 18mm IED has a good output characteristics. The fabricated dry-type surface myoelectric sensor is evaluated for the upper-limb amputee. |