Title |
A Study on Improvement of the Accuracy of SV Measurement obtained by Hand to Hand Impedance. |
Authors |
윤찬솔(Yoon, Chan-Sol) ; 염호준(Yeom, Ho-Jun) |
DOI |
https://doi.org/10.5370/KIEE.2015.64.8.1251 |
Keywords |
Impedance cardiography ; Hand to hand ; Stroke volume ; Cardiac output ; Non-invasive ; Cardiovascular |
Abstract |
The purpose of this study is to measurement the cardiac output using hand to hand impedance method to provide convenience to user when measuring SV(stroke volume) with the use of ICG(Impedance Cardiography). We suggest the optimized modified formula, which can be applied when using impedance with the use of hand to hand Impedance. To verify this formula, a SV from transthoracic approach and hand to hand approach are compared for the 36 subjects, respectively. The acquired data was analyzed by using LabVIEW 8.6, analysis was implemented by SPSS 12.0. Fine difference was shown by individual. We showed that as a result of analyzing the ICG measurement value on thoracic and hand to hand, the correlation with SV was r=0.716, thereby having indicated the results of regression model in relatively high correlation. |